2 add sample - properties – Metrohm viva 1.1 User Manual

Page 468

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

456

■■■■■■■■

viva 1.1

Identification

Description

CYL

Cylinder volume of the dosing unit used for the com-
mand

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

not valid (variable not available) = Command
has not yet been started

VOL

Currently dosed volume in L (Number)

5.5.2.4.4.2

ADD SAMPLE - Properties

Dialog window: Method

ADD SAMPLE Properties... ADD SAMPLE - 'Com-

mand name'

Command name

Name of the command.

Entry

25 characters

Addition
Selection of the addition mode for the sample.

Selection

Add manually | Already added | Add with dos-
ing device

Default value

Add manually

Add manually
Manual addition of the sample, at which time the track is stopped. A
message appears and an acoustic signal is heard.
Already added
The sample is located already in the measuring vessel or was channeled
into the measuring vessel, e.g. with a peristaltic pump. The track is not
stopped. No message appears and no acoustic signal is heard.
Add with dosing device
Automatic addition of the solution with a dosing device. The Sample
amount
is added for undiluted samples, the Analytical volume for
diluted ones. The sample is added through the Dosing port 1 defined
in the configuration for the dosing unit (see Chapter 6.6.3.2, page
785)
.

Advertising