7 Devices

viva 1.1

■■■■■■■■

949

Selection

not defined | Pump | Stirrer | Valve | Gen. out-
put
| Alarm signal

Default value

not defined

NOTE

A stepping motor can in addition have a port of the Type = Brake.

The port information for this port type is automatically read out of the
Stepping Motor Controller.

Set default value

Shows the default value which the output takes directly after starting and
exiting viva. The default value is also set after an error or when canceling
the method.

NOTE

Behavior at Emergency stop:

All digital outputs that are not of the Type = Alarm signal are set
to the default value when an emergency stop takes place.

A digital output of Type = Alarm signal is set to the non-default
value
value when an emergency stop takes place.

Digital outputs assigned to a digital input of the Type = Level
detector
or Leak detector do not change their status when an
emergency stop takes place.

Selection

active | inactive

Default value

inactive

Default values for commands

Parameter 1

The meaning of Parameter 1 depends on the Type set for the digital
output.

For Type = Stirrer: stirring performance in %, pulse duration dependent
on Parameter 2.

Input range

-214,783,647 - 214,783,647

Default value

100

Parameter 2

The meaning of Parameter 2 depends on the Type set for the digital
output.