Dosing
Input range
0.00000 - 99,999.9 mL
Default value
10.0000 mL
Aspirate
Input range
0.00000 - 50.0 mL
Default value
10.0000 mL
Rate
Speed with which the functions Dosing, Fill, Aspirate, Eject to stop,
Eject to end volume and Exchange position are carried out. The max-
imum speed depends on the cylinder volume of the dosing unit used. The
aspiration and expulsion of the sample should take place at a speed of <
10 mL/min.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
CAUTION
The maximum permitted dosing rate of 4 mL/min may not be exceeded
for capillaries with a diameter < 1 mm.
5.5.2.4.7
PREP
5.5.2.4.7.1
PREP - Overview
Dialog window: Method
▶ PREP ▶ Properties... ▶ PREP - 'Command name'
Command for rinsing the cylinder and tubings of a dosing unit.
Instruments
This command can be executed with the following instruments:
Voltammetry: 894
Dosing Interface: 846
Sample changer: 814, 815, 858, 919
Photometer: 089
Appearance
The command has the following appearance: