7 prep, 1 prep - overview, Prep – Metrohm viva 1.1 User Manual

Page 479

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1

■■■■■■■■

467

Dosing

Input range

0.00000 - 99,999.9 mL

Default value

10.0000 mL

Aspirate

Input range

0.00000 - 50.0 mL

Default value

10.0000 mL

Rate

Speed with which the functions Dosing, Fill, Aspirate, Eject to stop,
Eject to end volume and Exchange position are carried out. The max-
imum speed depends on the cylinder volume of the dosing unit used. The
aspiration and expulsion of the sample should take place at a speed of <
10 mL/min.

Input range

0.01 - 166.00 mL/min

Selection

maximum

Default value

maximum

CAUTION

The maximum permitted dosing rate of 4 mL/min may not be exceeded
for capillaries with a diameter < 1 mm.

5.5.2.4.7

PREP

5.5.2.4.7.1

PREP - Overview

Dialog window: Method

PREP Properties... PREP - 'Command name'

Command for rinsing the cylinder and tubings of a dosing unit.

Instruments
This command can be executed with the following instruments:

Voltammetry: 894

Dosing Interface: 846

Sample changer: 814, 815, 858, 919

Photometer: 089

Appearance
The command has the following appearance: