8 Manual control

viva 1.1

■■■■■■■■

967

p = set pulse (pulse length = 1 s. If a pulse with a different length is to be
output, then a corresponding template must be defined for this purpose.)

Sample changers
Selection

Bit pattern with exactly 14 characters (0, 1, *,
p)
| ************** | signal pattern

Default value

**************

894
Selection

Bit pattern with exactly 14 characters (0, 1, *,
p)
| ************** | signal pattern

Default value

**************

Spectrometer
Selection

Bit pattern with exactly 10 characters (0, 1, *,
p)
| ********** | signal pattern

Default value

**********

The output lines and bits are numbered from right to left:

Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0

Bit

13 12 11 10 9 8 7 6 5 4 3 2 1 0

Examples:

************1* sets the output line 1 to active (= set).

************0* sets the line to inactive.

NOTE

We recommend masking the irrelevant output lines with an asterisk *
so as not to modify these line statuses.

[Set]

Sets the bit pattern defined under Output signal.

8.4.5

Sampler changer functions

8.4.5.1

Sample changer - Overview

Program part: Manual control

If a tower of a sample changer is selected in the subwindow for the device
selection, then the functions and parameters that are possible with this
sample changer are displayed in the subwindow for functions/parameters.

Sample changer functions
The sample changer functions are displayed on the following tabs: