2 wait - properties – Metrohm viva 1.1 User Manual

Page 615

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1

■■■■■■■■

603

5.5.2.10.4.2

WAIT - Properties

Dialog window: Method

WAIT Properties... WAIT - 'Command name'

Command name

Name of the command.

Entry

25 characters

Wait
Selection of the following options:

Selection

Stop track and waiting for [Continue] | Stop
all tracks and waiting for [Continue]
| Waiting
time

Default value

Stop track and waiting for [Continue]

Stop track and waiting for [Continue]
If this option is selected, then the track in which the WAIT command
is located is stopped and a message is displayed. The method run can
be continued only by pressing [Continue].
Stop all tracks and waiting for [Continue]
If this option is selected, then all active tracks are stopped and a mes-
sage is displayed. The method run can be continued only by pressing
[Continue].
Waiting time
If this option is selected, then the method run will be continued auto-
matically after the waiting time expires. The message text defined
under Message appears during the waiting time.

Waiting time

This parameter can be edited only if Wait = Waiting time.

Input range

0 - 9,999.9

Default value

0

Selection

s | min

Default value

s

Message
The message defined here is displayed with tracks stopped or during the
waiting time. The text editor for entering or changing the message is

opened with

or by double-clicking on the text field (see Chapter

2.4.2, page 78). The formula editor can also be opened within the text
field (see Chapter 2.3, page 21).

Entry

Text (unlimited)