Metrohm viva 1.1 User Manual
Page 654
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
642
■■■■■■■■
viva 1.1
2 Plating time
Idle until the period defined in the Plating time parameter has
expired.
Sweep (measure open-circuit potential)
1 Measurement duration
The circuit is opened and the potential between the working elec-
trode and the reference electrode is measured for the indicated mea-
surement duration.
Post-treatment
1 Cleaning potential
Cleaning potential is applied to the electrodes.
2 Cleaning time
Idle until the period defined in the Cleaning time parameter has
expired.
only for hydrody-
namic measurement
= on
3 Stirrer off
The stirrer is switched off.
only for standby
potential = off
4 Potential off
The potential applied to the electrodes is switched off.
only for standby
potential
≠
off
5 Standby potential on
Standby potential is applied to the electrodes.
Parameters
The parameters for the CP command are set on the following tabs:
■
■
■
■
■