7 Devices

viva 1.1

■■■■■■■■

915

NOTE

Behavior at Emergency stop:

All digital outputs that are not of the Type = Alarm signal are set
to the default value when an emergency stop takes place.

A digital output of Type = Alarm signal is set to the non-default
value
value when an emergency stop takes place.

Digital outputs assigned to a digital input of the Type = Level
detector
or Leak detector do not change their status when an
emergency stop takes place.

Selection

active | inactive

Default value

inactive

Default values for commands

Parameter 1

The meaning of Parameter 1 depends on the Type set for the digital
output.

For Type = Stirrer: stirring performance in %, pulse duration dependent
on Parameter 2.

Input range

-214,783,647 - 214,783,647

Default value

100

Parameter 2

The meaning of Parameter 2 depends on the Type set for the digital
output.

For Type = Stirrer:

0: Duration of the switch-on pulse set to 500 ms.

>0: With a stirring performance of 100% (Parameter 1 = 100), pulse
duration 10 ms
Example:
Parameter 1
= 100, Parameter 2 = 30: pulse duration switched on
300 ms, switched off 0 ms
Parameter 1 = 75, Parameter 2 = 30: pulse duration switched on
225 ms, switched off 75 ms

Input range

-214,783,647 - 214,783,647

Default value

0