Metrohm viva 1.1 User Manual
Page 910
7.8 RS-232 device
■■■■■■■■■■■■■■■■■■■■■■
898
■■■■■■■■
viva 1.1
Stop bit
Number of stop bits.
Selection
1 | 2
Default value
1
Handshake
Type of the data transfer protocol.
Selection
None | HW | SW
Default value
None
Timeout
Maximum waiting time for receiving characters. If this time is exceeded,
receiving will be canceled.
Input range
1,000 - 9,999 ms
Default value
2,000 ms
Terminator for send
Terminator for send in hexadecimal code.
Selection
\0D | \0D\0A | \0D\0A\0A | \0D\0D\0A
Default value
\0D\0A
(\0D = CR, \0A = LF)
Terminator for receive
Terminator for receive in hexadecimal code.
Selection
\0D | \0D\0A | \0D\0A\0A | \0D\0D\0A
Default value
\0D\0A
(\0D = CR, \0A = LF)
Code page
Code page used for the data transfer.
Selection
Cp437 | Cp850 | Cp852 | ASCII
Default value
Cp437
Send characters one by one
on | off (Default value: off)
If this check box is activated, then each character is sent individually and
the waiting time has to elapse before the next character is sent.
Delay time
Waiting time until the next character is sent.