6.8 Colorimetric sensors subwindow

■■■■■■■■■■■■■■■■■■■■■■

806

■■■■■■■■

viva 1.1

Calibration data monitoring
on | off (Default value: off)

If this check box is activated, then the validity of the calibration is moni-
tored.

Calibration interval

Validity of the calibration in days. If a value is entered here, then the date
for the next calibration will be adapted automatically. The days are moni-
tored, i.e., the calibration becomes invalid when the date changes and not
at the time when the calibration took place.

Only editable for Calibration data monitoring = on, otherwise inactive.

Input range

0 - 999 days

Default value

999 days

Next calibration

Date on which the next calibration is to be carried out. The date can be

selected by clicking on

in the Select date dialog window. After the

date has been entered, the value for Calibration interval will be auto-
matically adjusted.

Only editable for Calibration data monitoring = on, otherwise inactive.

Default value Calibration date+ 999 days

Message

Message by e-mail

on | off (Default value: off)

The message is additionally sent to the address defined under [E-mail...]
if this check box is activated. The message is sent in text format.

Editable only if Calibration data monitoring = on, otherwise inactive.

[E-mail...]

[E-mail...] opens the Send e-mail window (see Chapter 2.5, page 79).

Active only if Message by e-mail = on.

Acoustic signal

on | off (Default value: off)

If this check box is activated, an acoustic signal will be emitted in addition
to the message.

Editable only if Calibration data monitoring = on, otherwise inactive.