5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

544

■■■■■■■■

viva 1.1

Input range

0 - 999,999 s

Default value

0 s

Max. waiting time

If the signal drift has not yet been reached, then the measured value will
be accepted when the maximum waiting time has elapsed. If the waiting
time has not been newly entered, then a waiting time that is suitable for
the drift will be calculated automatically according to the following for-
mula:

Input range

0 - 999,999 s

Default value

154 s

Measuring interval

Time interval for entering a measured value in the measuring point list.

Input range

0.25 - 999,999.00 s (Increment: 0.25)

Default value

2.00 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Input range

0 - 4,000 mAU

Selection

off

Default value

off

off
No stop.

Measurement without drift control
If the Measurement without drift control option has been selected,
then the following parameters will be displayed:

Max. waiting time

Maximum period of time for measurement.

Input range

0 - 999,999 s

Default value

120 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Input range

0.25 - 999,999.00 s (Increment: 0.25)

Default value

2.00 s