5 pump, 1 pump - overview, Pump – Metrohm viva 1.1 User Manual

Page 448

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

436

■■■■■■■■

viva 1.1

5.5.2.3.5

PUMP

5.5.2.3.5.1

PUMP - Overview

Dialog window: Method

PUMP Properties... PUMP - 'Command name'

Command for the control of the pumps that are connected to or
installed on the sample changer.

Instruments
This command can be executed with the following instruments:

Sample changer: 814, 815, 858, 919

Appearance
The command has the following appearance:

Parameters
The parameters for the PUMP command are configured in the following
dialog window:

PUMP - Properties

Command variables
The following command variables are generated by the PUMP command
in the method run and can be used in formulas under the designation
'Command name.Variable identification':

Identification

Description

BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

not valid (variable not available) = Command
has not yet been started

Advertising