5 pump, 1 pump - overview, Pump – Metrohm viva 1.1 User Manual
Page 448
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
436
■■■■■■■■
viva 1.1
5.5.2.3.5
PUMP
5.5.2.3.5.1
PUMP - Overview
Dialog window: Method
▶ PUMP ▶ Properties... ▶ PUMP - 'Command name'
Command for the control of the pumps that are connected to or
installed on the sample changer.
Instruments
This command can be executed with the following instruments:
Sample changer: 814, 815, 858, 919
Appearance
The command has the following appearance:
Parameters
The parameters for the PUMP command are configured in the following
dialog window:
■
Command variables
The following command variables are generated by the PUMP command
in the method run and can be used in formulas under the designation
'Command name.Variable identification':
Identification
Description
BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command
has never been started
FIN
Command status:
1 = Command has ended at least once
0 = Command has not yet ended
not valid (variable not available) = Command
has not yet been started