5 Method

viva 1.1

■■■■■■■■

549

Input range

0.1 - 600.0 s (Increment: 0.1)

Default value

1.0 s

Signal drift

The measurement is canceled if the value has fallen below the signal drift
defined here.

Input range

0.1 - 999.0 °C/min

Default value

0.1 °C/min

Min. waiting time

The signal drift is only checked after the minimum waiting time defined
here has been exceeded.

Input range

0 - 6,500 s

Default value

0 s

Max. waiting time

The measurement is canceled as soon as the maximum waiting time
defined here has been exceeded. If the waiting time has not been newly
entered, then a waiting time that is suitable for the drift will be calculated
automatically according to the following formula:

Input range

0 - 6,500 s

Default value

457 s

Stop measured value

The measurement is canceled when the specified measured value since
the start of the measurement is reached.

The measurement is also canceled if the Min. waiting time has not yet
been exceeded.

It will also be canceled if the value has not yet fallen below the Signal
drift
.

Input range

0.0 - 100.0 °C

Selection

off

Default value

off

off
No stop.

Measurement without drift control
If the Measurement without drift control option has been selected,
then the following parameters will be displayed: