2 meas t - general/hardware – Metrohm viva 1.1 User Manual

Page 558

Advertising
background image

5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

546

■■■■■■■■

viva 1.1

Command variables
The following command variables are generated by the command MEAS
T
in the method run and can be used in formulas under the designation
'Command name.Variable identification':

Identifi-
cation

Description

BSY

Command status:

1 = BUSY, HOLD or ERROR

0 = READY

invalid (variable not available) = Command has never
been started

DBL

Total duration for the processing of the command in s

ETE

End temperature (temperature after the command has
been processed) in °C

FIN

Command status:

1 = Command has ended at least once

0 = Command has not yet ended

invalid (variable not available) = Command has never
been started

ITE

Initial temperature (temperature before start conditions are
processed) in °C

NMP

Number of measuring points in measuring point list

STY

Type of stop with which the command was stopped: 1 =
finished normally
, 0 = stopped manually with [Quit]
or
[Stop]

5.5.2.7.7.2

MEAS T - General/Hardware

Tab: Method

MEAS T Properties... General/Hardware

Command name

Name of the command.

Entry

1 - 25 characters

The general parameters for the instrument, the sensor and the stirrer are
defined on this tab.