5.5 Method run subwindow

■■■■■■■■■■■■■■■■■■■■■■

484

■■■■■■■■

viva 1.1

Selection

Linear regression | Quadratic regression |
cubic | automatic

Default value

Linear regression

Confidence interval

Range within which the measured values for the determination of the cali-
bration curve must lie. Values outside this range are called outliers; they
are not integrated in the calculation of the calibration coefficient.

Selection

99 | 98 | 95 | off

Default value

98

off
The confidence interval will not be checked.

5.5.2.5.4

CAL MEAS Opt

5.5.2.5.4.1

CAL MEAS Opt - Overview

Dialog window: Method

CAL MEAS Opt Properties... CAL MEAS Opt - 'Com-

mand name'

Command for measuring standard solutions for the calibration of
colorimetric sensors
, used together with the CAL LOOP Opt com-
mand.

Instruments
This command can be executed with the following instruments:

Spectrometer: Avantes

Photometer: 089

Appearance
The command has the following appearance:

Parameters
The parameters for the CAL MEAS Opt command are set on the follow-
ing tabs:

General/Hardware
Selection of instrument type and colorimetric sensor.

Measuring parameters
Parameters for setting the measurement procedure.