7 Devices

viva 1.1

■■■■■■■■

891

FPGA version

Shows the FPGA version of the device.

Set to work

Shows the date on which the device was automatically added to the
device table.

Remarks

Remarks about the device.

Entry

1,000 characters

Default value

'empty'

7.7.3

Properties - Settings

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Settings

Light source

Wavelength

Wavelength which has been set on the Photometer. If a value is defined
here, a check is performed for each command containing a specified
wavelength whether the wavelength specified in the command corre-
sponds with the wavelength defined here for the device.

Input range

100 - 2,000 nm

Selection

'empty'

Default value

'empty'

Light source always active

on | off (Default value: on)

If this check box is activated, then the LED on the Photometer is always
on. Otherwise, the firmware will switch it on before the measurement and
switch it off after the measurement.

Heater

Initial temperature

Initial temperature of the device (temperature of the cuvette). Entering an
initial temperature means that the heater will heat up to the set tempera-
ture when the device is switched on. off means that the heater will not be
switched on.

Input range

20 - 60 °C

Default value

30 °C

Selection

off