Metrohm viva 1.1 User Manual

Page 43

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

2 General program functions

viva 1.1

■■■■■■■■

31

NOTE

In the case of variables with the index {x}, the required number 1 - 9
must be entered for x (e.g., RS.CVS 3.VAR{3}.VTOT for the total vol-
ume in the measuring vessel after the third addition).

If no index is specified, then the last index will be used automatically
(e.g., RS.CVS 3.VAR.VTOT for the total volume after the last addi-
tion).

Identifica-
tion

Description

Commands

ACO

Addition counter = current number of completed runs of the command

ADD SAMPLE
DT, ADD STD

BP{x}.CNT

Intensity for the break point x (1 - 9) in counts

MEAS Opt

BP{x}.MEA

Measured value for the break point x (1 - 9) in the unit of the measured value

MEAS Opt

BP{x}.TEM

Temperature for the break point x (1 - 9) in °C

MEAS Opt

BP{x}.TIM

Time for the break point x (1 - 9) in s

MEAS Opt

BP{x}.TRN

Transmission for the break point x (1 - 9) in %

MEAS Opt

BSY

Command status:

1 = BUSY, HOLD or ERROR

0 = READY

invalid (variable not available) = Command has never been started

All except
START and
END

C0

Coefficient c0 of the calibration curve

CAL LOOP
Opt, CAL Spec

C1

Coefficient c1 of the calibration curve

CAL LOOP
Opt, CAL Spec

C2

Coefficient c2 of the calibration curve

CAL LOOP
Opt, CAL Spec

C3

Coefficient c3 of the calibration curve

CAL LOOP
Opt, CAL Spec

CAL

Status of the calibration: 1 = calibrated normally, 0 = canceled, because ref-
erence wavelength was not found.

CAL Spec

COD

Coefficient of determination (R

2

), which is calculated on the basis of the calibra-

tion function and of the confidence interval.

CAL LOOP Opt

CP{#}.MEA

Measured wavelength of the calibration point in nm. If no measured peak can be
assigned, then not found will be entered.

CAL Spec

CP{#}.REF

Reference wavelength of the calibration point in nm.

CAL Spec