3 properties - settings, 4 properties - detector, Properties - settings – Metrohm viva 1.1 User Manual

Page 896: Properties - detector, Settings, Detector

Advertising
background image

7.6 Avantes Spectrometer

■■■■■■■■■■■■■■■■■■■■■■

884

■■■■■■■■

viva 1.1

7.6.3

Properties - Settings

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Settings

Light source

Type

Type of light source used.

Selection

continuous | pulsed

Default value

continuous

Pulses per integration time

Frequency at which the lamp flashes. This parameter can be edited only if
Type = pulsed.

Input range

0 - 1,000

Default value

10

Analog output

Analog output

on | off (Default value: on)

If this check box is activated, then the measuring signal is output at the
analog output of the spectrometer. If the connected spectrometer is not
equipped with an analog output, then this setting will be ignored.

Sensitivity

Input range

1 - 214,783,647 mAU/mV

Default value

100 mAU/mV

7.6.4

Properties - Detector

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Detector

The Detector tab shows information on the detector that is built into the
spectrometer.

Detector type

Shows the detector type.

Minimum wavelength

Shows the minimum wavelength of the measurable range in nm.

Maximum wavelength

Shows the maximum wavelength of the measurable range in nm.