7.10 IO controller

■■■■■■■■■■■■■■■■■■■■■■

936

■■■■■■■■

viva 1.1

Selection

'Port designation' | calculated value

'Port designation'
All analog input ports can be selected.

0% Mapping

Measured value to be used when the Low level (0 %) is applied as input
value.

Only editable for Control variable = calculated value, otherwise
grayed out and not available for editing.

Input range

-214,783,647.000 - 214,783,647.000

Default value

0 or controlled variable setting, respectively

100% Mapping

Measured value to be used when the High level (100 %) is applied as
input value.

Only editable for Control variable = calculated value, otherwise
grayed out and not available for editing.

Input range

-214,783,647.000 - 214,783,647.000

Default value

100 or controlled variable setting, respec-
tively

Unit

Unit of the measured value to be used.

Only editable for Control variable = calculated value, otherwise
grayed out and not available for editing.

Entry

50 characters

Default value

% or controlled variable unit, respectively

Default value

Selection or input of the standard value which is to be used for the input if
the result is invalid.

Input range

0% mapping of the input - 100% mapping of
the input

Selection

last value | Setpoint value

Default value

last value

Setpoint value

Setpoint value to which the control is to be set.