7 special results for dt, mlat, lat and rc, 1 standardized area – Metrohm viva 1.1 User Manual

Page 727

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

viva 1.1

■■■■■■■■

715

5.6.7.6.3

Absolute standard deviation of the concentration of the substance
in the sample

Variable name

RS.’Command name’.’Substance name’.CONC.ASD

Description

Absolute standard deviation of the concentration of the substance in the
sample.

Decimal places

0…2…6

Unit

As for concentration of the substance in the sample (see Chapter
5.6.7.6.1, page 713)

5.6.7.6.4

Relative standard deviation of the concentration of the substance in
the sample

Variable name

RS.’Command name’.’Substance name’.CONC.RSD

Description

Relative standard deviation of the concentration of the substance in the
sample.

Decimal places

0…1…6

Unit

%

5.6.7.7

Special results for DT, MLAT, LAT and RC

5.6.7.7.1

Standardized area

Variable name

RS.'Command name'.'Substance name'.VAR{x}.REP{y}.AREA-
NORM
(if the command is called by CALL VA)

Calibration method DT:

RS.'Command name'.'Substance name'.VMS.REP{y}.AREANORM
(if the command is called by CALL VMS)

Calibration method RC:

RS.'Command name'.'Substance name'.ELECTRO-
LYTE.REP{y}.AREANORM
(if the command is called by CALL ELEC-
TROLYTE
)

Description

Standardized area of a peak.

Data type

Number

Decimal places

0…2…6

Unit

Remark

Only for Calibration method = DT, RC.

Calculation

Advertising