3 swing, 1 swing - overview, Swing – Metrohm viva 1.1 User Manual
Page 442
5.5 Method run subwindow
■■■■■■■■■■■■■■■■■■■■■■
430
■■■■■■■■
viva 1.1
Shift direction
Selection of the shift direction.
Selection
auto | + | –
Default value
auto
auto
The shift direction with the smallest rotation angle is selected automat-
ically.
+
Rack rotates counterclockwise.
–
Rack rotates clockwise.
Swing rate
Selects the rate of the robotic arm when moving to a rack position or a
special beaker (only with multiple row racks).
Input range
10 - 55 °/s
Default value
55 °/s
5.5.2.3.3
SWING
5.5.2.3.3.1
SWING - Overview
Dialog window: Method
▶ SWING ▶ Properties... ▶ SWING - 'Command name'
Command for swinging the robotic arm. This command cannot be
executed unless the Sample Processor is equipped with a 786 Swing
Head with robotic arm.
Instruments
This command can be executed with the following instruments:
Sample changer: 814, 815, 858, 919
Appearance
The command has the following appearance:
Parameters
The parameters for the SWING command are configured in the following
dialog window:
■