5 Method
viva 1.1
■■■■■■■■
491
Measurement without drift control
If the Measurement without drift control option has been selected,
then the following parameters will be displayed:
Max. waiting time
Maximum period of time for measurement.
Input range
0 - 999,999 s
Default value
120 s
Measuring interval
Time interval for entering a measured value in the measuring point list.
Avantes Spectrometer
Input range
0.1 - 999,999.0 s (Increment: 0.1)
Default value
2.0 s
089 Photometer
Input range
0.25 - 999,999.00 s (Increment: 0.25)
Default value
2.0 s
Stop measured value
Stop when the entered measured value has been reached since the start
of the measurement.
Input range
0 - 4,000 mAU
Selection
off
Default value
off
off
No stop.
Temperature
This section is displayed only when Device type = 089 Photometer.
Temperature
Measuring temperature, which can be entered manually. The temperature
is measured continuously when Temperature measurement = contin-
uous (tab General/Hardware).
Input range
20.0 - 60 °C
Default value
25.0 °C