Bio-Plex Manager Software 6.1 User Guide | Preparing Protocols

88

NOTE: Each external standard you select must match the region of a selected
analyte in the protocol. For example, if you selected analytes in regions 19, 32,
52, and 73 in the Select Analytes window (page 59), you can select only
external standards in those same regions. (Note that the analyte name does
not have to match, only the region.) Also, each analyte must come from a
different region.

When you are finished, the Selected External Standards list should include all
the analytes that you want to use as external standards, and their source
file(s).

Figure 78. Selected external standards

Click the External Standards Info tab to enter additional information about
your external standards.