Bio-Rad Bio-Plex Software® Upgrades and Conversions User Manual

Page 47

Advertising
background image

Calibration

35

To open the Calibration Log, go to the View menu and select Calibration Log.
The Calibration Log viewer opens.

Figure 21. Calibration Log viewer

Each calibration is listed by date and time. If you are using Bio-Plex Manager
Security Edition in Secure Mode, the User and Access Level columns list
information about the user who was logged into the application when each
calibration was performed. The Result column notes the result of each
calibration, including whether it passed or failed.

NOTE: If you see a drastic change in a detector's voltage from one calibration
to the next, it could indicate a problem with the instrument. A steadily
increasing detector voltage may indicate that the laser is decreasing in
intensity.

On the toolbar, click Calibration Detail

to display additional information

for each calibration, including bead count, system temperature and pressure,

and the reader software and firmware versions. Click Calibration Log

to

return to the default view.

For CAL1-only calibrations, the RP1 column displays an entry of NA (not
applicable). For CAL2-only calibrations, the DD, CL1, and CL2 columns
display entries of NA.