Raw data table error codes, Printing the raw data, Raw data table error codes printing the raw data – Bio-Rad Bio-Plex Software® Upgrades and Conversions User Manual
Page 164
Bio-Plex Manager Software 6.1 User Guide | Analyzing the Results
152
Raw Data Table Error Codes
If data cannot be measured or a sampling error occurs in a well, it will be
flagged in the Raw Data table. The possible error conditions in the table are
listed below.
Printing the Raw Data
You can print the Raw Data table using the Print command
on the main
toolbar or File menu. Print Preview on the File menu displays the table as it will
appear in the printout.
Column
Error Code
Cause
F1
***
No data present
Sampling
Errors
1
Low bead number detected in the well
2
Aggregated beads detected in the well
3
Classify efficiency problem detected in the well
4
Region selection problem detected in the well
5
Platform temperature problem detected in the well