Raw data table error codes, Printing the raw data, Raw data table error codes printing the raw data – Bio-Rad Bio-Plex Software® Upgrades and Conversions User Manual

Page 164

Advertising
background image

Bio-Plex Manager Software 6.1 User Guide | Analyzing the Results

152

Raw Data Table Error Codes

If data cannot be measured or a sampling error occurs in a well, it will be
flagged in the Raw Data table. The possible error conditions in the table are
listed below.

Printing the Raw Data

You can print the Raw Data table using the Print command

on the main

toolbar or File menu. Print Preview on the File menu displays the table as it will
appear in the printout.

Column

Error Code

Cause

F1

***

No data present

Sampling
Errors

1

Low bead number detected in the well

2

Aggregated beads detected in the well

3

Classify efficiency problem detected in the well

4

Region selection problem detected in the well

5

Platform temperature problem detected in the well

BioPlex_6.book Page 152 Friday, September 23, 2011 11:55 AM