Bio-Plex Manager Software 6.1 User Guide | Analyzing the Results

148

Viewing/Changing the Protocol Settings

In the Results window, you can review and in some cases change the plate
formatting, analyte selections, sample concentrations, etc., from the original
Protocol file to display different data or recalculate data in the Results file
tables. These controls are located in the toolbar across the top of the Results
window.

NOTE: Any changes you make in the Results file will not change the original
Protocol file or the data from the reading. They will change only how these
data are displayed in the Results file tables.

Click the Describe Protocol button

in the Results toolbar to review or

change the description of the Protocol, as described on page 59.

Click the Select Analytes button

in the Results toolbar to review the

analyte selection or select different analytes, as described on page 59.

NOTE: During a reading, the array reader always detects all the analytes in the
sample, including any you have not selected. However, analytes that were
detected, but not selected, will not appear in the final reports and tables. After
a reading, you can go back and change your analyte selections, and any
detected analytes will appear in the tables.

Click the Format Plate button

in the Results toolbar to review or change

the formatting of the plate. Plate formatting is described on page 68.

NOTE: Only formatted wells are read by the array reader. You must define all
the wells you want to read before running a Protocol. However, you can
change the formatting of a defined well (for example, from unknown to
standard) after a reading, and the analysis of that well will change.

Click the Enter Controls Info button

in the Results toolbar to review or

change the concentrations of the controls, as described on page 96.

Click the Enter Sample Info button

in the Results toolbar to review or

change the sample information, as described on page 98.

BioPlex_6.book Page 148 Friday, September 23, 2011 11:55 AM