Bio-Plex Manager Software 6.1 User Guide | Controlling the System

26

Bio-Plex MCV Plate IV

Bio-Plex Manager 5.0 and later requires the use of the Bio-Plex MCV
(Maintenance, Calibration and Validation) Plate IV. This plate contains wells
marked for the different types of fluids used in validation, washing, calibration,
and other functions. The Bio-Plex MCV Plate IV has been modified from the
previous MCV plate to work with Bio-Plex Validation Kit 4.0 and its Reporter
and Classify bead sets. It also includes two open needle wells for adjusting
needle heights for microplates or PCR plates.

NOTE: Bio-Plex Manager 5.0 and later require the use of the Bio-Plex MCV
Plate IV, the Bio-Plex Validation Kit 4.0, and the Bio-Plex Calibration Kit.

Figure 12. Bio-Plex MCV Plate IV

When a particular procedure such as calibration requires you to add solutions
to the Bio-Plex MCV Plate IV, the dialog box describing the procedure
includes a diagram of the MCV Plate IV with the wells to be loaded highlighted
in blinking yellow.

Figure 13. Highlighted wells in diagram indicate wells to be filled

Wells highlighted

in yellow