The following table is based on the selections made in Normalization Options.
It also describes how the Norm (Ratio) is calculated:

NOTES:

Replicates are done at the plate level. See the section on plates to
define replicates

When using multiple replicates, the Norm (Ratio) for each replicate
is calculated as described in Case A, B, C, D, and E. For each
sample type, the replicates' average is the Norm (Ratio)

For information about the formulas Bio-Plex Manager uses to
calculate normalization values, see Normalization Formulas on
page 223

Case

Internal
Controls

Control
Sample

How Norm (Ratio) is Calculated

A

Single

None

Ratio of (FI-Bkgd) for the analyte and the
housekeeping gene:

Norm (Ratio) = (FI-Bkgd)/(FIhk-Bkgdhk)

B

Multiple

None

Geometric mean of the normalized (FI - Bkgd)
ratios for each housekeeping gene

C

Single

Single

Normalized (FI - Bkgd) ratio/(FI - Bkgd) control
ratio for the control (as defined in Case A):

Norm (Ratio) = Norm (Ratio) single gene/Norm
(Ratio) control

D

Multiple

Single

(Ratio) multiple internal control divided by the
Norm (Ratio) multiple control (as defined in
Case B) for the control:

Norm (Ratio) = Norm (Ratio) multiple internal/
Norm (Ratio) multiple…

E

None

Single

Ratio of (FI-Bkgd) for the analyte and the
control:

Norm (Ratio) = (FI-Bkgd)/(FI control-
Bkgdcontrol)