Report Table Error Codes
Certain data that cannot be measured or calculated will be flagged in the
Report Table.
Columns and their possible error codes are listed below.
Column
Error Code Cause
FI
***
No data present
---
Marked as an outlier
FI - Background
***
No data present
(negative values, if any will be displayed
---
Marked as an outlier
SD
***
No data present
CV%
***
No data present
Obs Conc (4PL/
5PL)
***
No data present
OOR>
Out of Range - Above asymptote of equation
OOR<
Out of Range - Below asymptote of equation
OOR
Out of Range - Math Error
*Value
Above or below standards FI
Obs Conc (Linear)
***
No data present
OOR
Negative extrapolated data
*Value
Positive extrapolated data or
Data above/below the standards FI
Exp Conc
***
No data entry - this value is input by user
OBS/Exp * 100
***
OOR displayed in Obs Conc column
Conc in Range
OOR< or
OOR>
Obs Conc is not within the specified recovery range
of standards conc
(the recovery range is specified by the user)
Sampling Error
1
Low bead number detected in the well
2
Aggregated beads detected in the well
3
Classify efficiency problem detected in the well
4
Region selection problem detected in the well
5
Platform temperature problem detected in the well