Bio-Plex Manager Software 6.1 User Guide | Starting Bio-Plex Manager

18

If you do not want to automatically connect to the array reader and platform
when you open the software, select the Do not connect at startup checkbox.
This is useful if you are not connected to an instrument or want to use the
software without connecting to the instrument. To re-enable automatic
connection at startup, select Reconnect from the Instrument menu (see next
section).

Disconnecting and Reconnecting

Communication between the computer and the array reader and platform is
closed when you exit Bio-Plex Manager. If you need to disconnect from the
array reader and platform without shutting down Bio-Plex Manager software
(for example, to troubleshoot the instrument), go to the Instrument menu and
select Disconnect.

To reconnect to the array reader and platform while Bio-Plex Manager is
running, select Reconnect from the Instrument menu. This dialog also enables
automatic connection each time you start the software, if it has been disabled
(see previous section).