Start up – Bio-Rad Bio-Plex Software® Upgrades and Conversions User Manual

Page 39

Advertising
background image

Start Up

27

Start Up

Startup is a series of fluidic functions that prepares the array reader to acquire
data. This process requires the Bio-Plex MCV Plate IV, distilled water, and
70% isopropanol. Startup takes approximately 10 minutes.

NOTE: The startup procedure can be performed while the optics are warming
up, but no validations, calibrations or runs should be performed before the
optics are warmed up. This takes an estimated time of 30 minutes.

Click the Start Up button

on the main toolbar or select the command

from the Instrument menu. Follow the step-by-step directions in the dialog
box for preparing the MCV Plate IV.

Figure 14. Start Up dialog

Insert the prepared plate into the microplate platform and click OK to begin
the startup process.

BioPlex_6.book Page 27 Friday, September 23, 2011 11:55 AM

Advertising