Bio-Plex Manager Software 6.1 User Guide | Preparing Protocols

72

C

HANGING

A

R

EPLICATE

G

ROUP

To change the replicate group of a well or group of wells:

1. Click the Select tool

, and click or drag the wells you want to change.

The wells appear highlighted.

2. Right-click the selection, and select Set Replicate Group from the

context menu.

3. In the pop-up box, select the desired replicate group number from the

pulldown list.

4. Click OK. The selection is changed to the specified group number.

Figure 60. Selecting a different replicate group

R

ESEQUENCING

W

ELL

N

UMBERS

If your well numbering is out of sequence and you want to correct it, select
Resequence Well Names from the Format Options menu. All the wells in the
plate are renumbered sequentially, based on whether you have selected
Autofill Across or Autofill Down. This is a cosmetic change, and does not
affect the analysis in any way.

D

EFINING

S

TANDARD

W

ELLS

Standard wells contain analytes of known concentration, which are used to
generate a standard curve of fluorescence intensity versus analyte
concentration. The regression equation for the curve is then used to calculate
the concentrations of your unknowns.