Multi-pattern configuration, Restore < test application > defaults – EXFO FTB-800 Series for FTB-1 User Manual

Page 116

Advertising
background image

Test Setup - Test Configurator, Timer, and System

106

FTB-700G/800 Series

BERT

Multi-Pattern Configuration

Note: Multi-Pattern Configuration is only available for DSN/PDH BERT test

application with DS1 signal (DS0 disabled) and when the pattern is
Multi-Pattern.

This feature allows sending each pattern for a specific duration,
sequentially and continuously.

Pattern represents a sequence of patterns that will be generated: 1111,
1in8, 2in8, 3in24, and QRSS.

Enable allows enabling the generation of each pattern individually in
the pattern sequence. All patterns are enabled by default. All 1’s (1111)
is disabled for Unframed test.

Individual Pattern Duration specifies the transmission duration of
each pattern: 15 s, 30 s, 45 s, 1 min, 2 min, 3 min (default),... up to
15 min.

Restore < Test Application > Defaults

Reverts the current test application to its default factory settings.