Clock, Clock synchronization, Ext clock in – EXFO FTB-800 Series for FTB-1 User Manual

Page 120

Advertising
background image

Test Setup - Test Configurator, Timer, and System

110

FTB-700G/800 Series

Clock

Clock

Allows the configuration of the clock synchronization.

From the Test menu, tap Setup, Test Configurator, and tap on the clock
block.

Clock Synchronization

Clock Mode allows to select the source clock that will be used for
transmission (TX).

Internal: Internal clock of the unit (STRATUM 3).

External: Clock signal from the EXT CLK port.

Ext Clock In

Note: Available when the Clock Mode is set to External.

Allows to set the external clock for test synchronization.

Interface Type allows the selection of the clock interface:
DS1 (default), E1, 2MHz, or 1PPS. 1PPS is automatically selected for
Dual Test Set in One-Way Latency measurement mode.

Ext Clock In on a green background indicates that a valid clock is
received on the EXT CLK port

LOS on a red background indicates that the clock is not valid on the
EXT CLK port.