Parameters, Table – EXFO FTB-800 Series for FTB-1 User Manual

Page 129

Advertising
background image

Test Setup - Test Configurator, Timer, and System

NetBlazer

119

EtherSAM - Burst

Parameters

Number of Burst Sequence represents the number of times, 1 to 100
(default is 2), the burst sequence will be repeated for the CBS and EBS
tests.

Refill Delay Ratio (%) represents the percentage of time required to
refill the CBS/EBS token buckets. The refill delay ratio is used for
post-burst delay and the remaining percentage is applied to the
pre-burst delay. The Refill Delay Ratio is configurable from 0 to 100%
(50% is the default as well as the standard’s minimum recommended
value).

Burst/IR Frame Ratio (%) represents the percentage of frames
transmitted at CIR rate for the CBS test and the percentage of frames
transmitted at CIR+EIR rate for the EBS test. The Burst/IR Frame
Ratio
is configurable from 10 to 90% (90% is the default as well as the
standard’s recommended percentage value).

Table

Note: Test time values are only displayed for enabled services.

Service No indicates the service number.

Service Name indicates the name of the service.

Direction, available with Dual Test Set, indicates results from local to
remote direction (L -> R) and remote to local direction (R -> L).

CBS Test Time (s) indicates the total time required to perform all burst
sequence iterations of the CBS test for this service.

EBS Test Time (s) indicates the total time required to perform all burst
sequence iterations of the EBS test for this service.

Total Burst Test Time (s) indicates the total time required to perform
all burst sequence iterations of both CBS and EBS tests for this service.