13 Suspend and Resume

Suspend and Resume allow respectively the FTB-1 and its running
applications to stop (Suspend) and to be quickly re-initialized (Resume)
when the unit is turned back on.

Suspend Mode

To enter the suspend mode, hold down the FTB-1 power button a few
seconds (approximately 1.5 seconds) until the unit beeps once, and select
either the Stand By or Hibernate button. In Suspend Mode, the NetBlazer
module is directly turned off and its configuration and injection information
are kept. The Suspend Mode remains active as long as battery power or AC
power is available. If the battery is drained the information is lost and the
Resume operation is no more possible.

The following conditions prevent the activation of the Suspend Mode:

A firmware download is in-progress

Resume operation is in-progress.

Note: The test will be stopped when entering the suspend mode meaning that the

test logger content will be lost.