Test Results

308

FTB-700G/800 Series

Logger and Alarms/Errors Logger

Details: Provides contextual information including the pass/fail verdict.

The following table displays the nature of information reported by type of
event:

Note: The Logger table can display up to 500 event entries. Once the Logger table

reports 500 event entries, a log full indicator appears and no further entries
is possible. However, the events in the Pending state will be updated if the
test is still running.

The Event Logger information will be cleared when:

the test is reset or started.

the unit is in suspended mode.

stopping the current test and navigating to other tests

the unit is restarted.

Note: An entry event remains in the Pending state as long as the event is not

completed and it is highlighted on a yellow background color.

Note: The Threshold Crossing events are displayed in red text color.

Type of Event

Nature of Information

Test Started

Start Date

Test Stopped

Pass/Fail Verdict

Alarm Events

Count value

Bit Error Events

Current Count and Total Count

SDT Events

Service Disruption Time

BER Threshold Crossing Event Value at the end of the test