INFICON Explorer Portable Gas Chromatograph User Manual

Page 114

Advertising
background image

SiteChart

Software

112

17. Ensure all compounds are correctly identified and then use these analyses to

calibrate

Explorer.

Interference Compounds

Compounds can be added to the library for identification purposes only. Using this
feature will allow compounds to be identified in a peak report but no concentration
data will be calculated. Interference compounds, included in some assays, are
provided in this way. Compounds listed in the library without response data still
need to be calibrated for retention time drift.

To assign interference compounds:

1. Store the compounds as outlined in Adding Compounds to the Library on page

162.

2. Select the compound in the library.

3. Select the point in the area/concentration table and click

Erase

.

The response data for the compound will be removed. The (0, 0) point and the
retention time information remain in the library.

4. Save the assay.