Maintenance and Troubleshooting

260

3. In SiteChart, select the Assay tab.

4. Open the assay and click

Send to

Explorer

.

Once the file has been downloaded, allow at least 30 minutes for

Explorer to

reach the assay set points.

5. Select the first column/detector pair.

If your application uses 2 or 3 columns you must repeat this procedure for each
column.
If you are using the ECD, you must repeat this procedure for each ECD library.

6. Run a chromatogram on the selected column/detector pair.

The chromatogram must contain all the compounds listed in your library.

7. Select the

Assay

tab, and select

Open

from the File menu to open the

chromatogram that was just run.

8. Select

Re-integrate

from the Analyzer menu to re-integrate the chromatogram.

Compounds matching library information will be identified.

9. Click on the first compound in the library.

10. Review the data in the library information box.

Compounds present in the calibration gas must be selected for calibration.
Compounds that are not present in the calibration gas must be removed from
the library.

NOTE: All compounds contained in the library must be contained in the calibration gas.

11. Ensure you have entered the correct calibration concentration.

This is the concentration of the compound in your calibration gas.
Blue dashed lines will be displayed where SiteChart is expecting to find the
compound.