Initiating Remote Communications

1. Enter the phone number at which the

Explorer and its modem are connected.

You can type parentheses and dashes, but they are not required. Type commas
to provide any delays required before the number is processed by the telephone
system.
If you are using a Hayes

TM

compatible modem with the default settings, each

comma inserts a delay of approximately 2 seconds.
If you are not using a Hayes

TM

compatible modem, the delays and the

characters that create the delays may be different. Refer to the modem user’s
manual.

2. Press the Dial button to initiate communication.

3. Monitor the modem status in the Dial Out dialog box. When the modem has

successfully connected with

Explorer, you can close the dialog box.

4. Select the Instrument tab and check the instrument status box.

The current values from

Explorer will be displayed.

You can now control the instrument from the computer. You can turn

Explorer

off from the computer but you cannot turn it on again.

Explorer must be powered

up from the local keypad by pressing the On/Off key.

Terminating Communication

When you are ready to terminate communication with

Explorer:

Open the Communication menu and select

Hang up

.

SiteChart will issue the command to the modem and disconnect the modem
from the telephone line.
You can also click

Hang Up

in the Dial Out dialog box. You can see the

modem status messages in this dialog box.

You can turn

Explorer off from the computer but you cannot turn it on again.

Explorer must be powered up from the local keypad by pressing the On/Off key.