Assay 1 standard method – INFICON Explorer Portable Gas Chromatograph User Manual
Page 288
Assay 1 User Notes
286
Assay 1 Standard Method
The Assay 1 standard method is installed on your
Explorer and this method includes
settings for each column. The table below provides typical settings, the exact values
will be specific to your instrument.
Column A
Column B
Column C
Method
Analysis Time (S)
3200
2100
1000
Oven Temperature ( C) 60
60
60
ECD Oven Temp. ( C)
60
60
60
Pressure (psi)
8
8
8
PID State
High Sensitivity High Sensitivity High Sensitivity
ECD State (if installed) On
On
On
Injection Method
Inject Method
Loop Injection
Loop Injection
Loop Injection
Inject
Time (S)
2
2
2
Pump Time (S)
20
20
20
Backflush Time (S)
1810
480
115
PID Integration Method
Automatic Integration
Disabled
Disabled
Disabled
Slope Up (mV/S)
0.5
0.5
0.3
Slope Down (mV/S)
1.0
1.0
1.0
Min. Height (mV)
0.0
0.0
0.0
Min. Area (mV S)
0.0
0.0
0.0
Filter Level
3
3
3
Integration Delay (S)
400
250
100