Assay 1 User Notes

286

Assay 1 Standard Method

The Assay 1 standard method is installed on your

Explorer and this method includes

settings for each column. The table below provides typical settings, the exact values
will be specific to your instrument.

Column A

Column B

Column C

Method
Analysis Time (S)

3200

2100

1000

Oven Temperature ( C) 60

60

60

ECD Oven Temp. ( C)

60

60

60

Pressure (psi)

8

8

8

PID State

High Sensitivity High Sensitivity High Sensitivity

ECD State (if installed) On

On

On

Injection Method
Inject Method

Loop Injection

Loop Injection

Loop Injection

Inject

Time (S)

2

2

2

Pump Time (S)

20

20

20

Backflush Time (S)

1810

480

115

PID Integration Method
Automatic Integration

Disabled

Disabled

Disabled

Slope Up (mV/S)

0.5

0.5

0.3

Slope Down (mV/S)

1.0

1.0

1.0

Min. Height (mV)

0.0

0.0

0.0

Min. Area (mV S)

0.0

0.0

0.0

Filter Level

3

3

3

Integration Delay (S)

400

250

100