Explorer Operator Manual

253

Figure 17 ECD Vent

5. Place the cotton swab in the container provided in the wipe-test kit and include

a data sheet stating that the wipe test was pe

erformed on a ECD, the serial

number of the instrument and the date of the test.

6. Replace the ECD access hatch.

7. Return the envelope to the leak testing center specified in the leak test kit and

request that a new wipe-test kit be sent with the test results.

WARNING

If the ECD does not pass the wipe test, contact

INFICON

immediately.