Explorer Operator Manual

151

Integration Delay Delays the start of the timer. Any peak eluting

before the timer starts will not be included in the
peak report.

Filter Level

Determines the amount of filtering to apply to the
signal. Filter 1 provides the most filtering.

Column

The

Column

command allows you to select the column for the next run on

Explorer. The selected column will remain enabled until it is de-selected. This

command over-rides the column sampling sequence on the instrument and forces
the run to be performed on the column selected here. The run will be performed
using the chromatographic settings for the selected column as set by the current
assay.

Start and Stop

The

Start

command starts a

Explorer run. Before beginning a run, ensure Explorer

has been set up correctly. Carrier gas must be supplied to the instrument.

To stop the analysis before it is complete, use the

Stop

command. If you stopped the

analysis after the sample was injected onto the column you must wait for the
duration of the analysis time for the column to be flushed before beginning another
analysis on the same column. When you stop an analysis,

Explorer will begin to

flush the column and move to the next column.

Calibrate

The

Calibrate

command calibrates

Explorer using the last analysis.

To calibrate the instrument:

1. Perform a run on the first column required by the assay.

The run must include all library compounds specific to the selected column.
Refer to the assay notes supplied with your assay for details.