INFICON Explorer Portable Gas Chromatograph User Manual

Page 152

Advertising
background image

SiteChart

Software

150

If you have selected a preset number of runs to be performed, SiteChart will
turn off the auto cycling option when it has completed the runs. If you are
cycling

Explorer continuously, you must turn auto cycling off from the

Analyzer menu.

9. In the Analyzer menu, select

Setup

. De-select the Auto Cycle check box and

click

OK

.

Auto cycling will stop when the analysis in progress is complete.

Integration Method

The options in the Integration Method dialog box allow you to view or alter the
integration method. After you change any of the integration parameters, the
chromatogram will automatically be re-integrated so you can see the effect of the
new integration values. The Integration Method is used to determine the compound
peaks and calculate the peak area. All of these parameters are set as part of the
method. Refer to Integration Method on page

175 and the assay notes before

making changes to the integration method.

When

Automatic Integration

is enabled the software automatically selects the

Slope Up and Slope Down values and minimum height based on a sampling of the
signal at the start of a run. Ensure

Automatic Integration

is selected.

Parameter

Description

Slope Up

The minimum slope required to start a peak.

Slope Down

The minimum slope required to stop a peak.

Minimum Height Any peak with a height less than the minimum

height will not be included in the peak report.

Minimum Area

Any peak with an area less than the minimum area
will not be included in the peak report.