Peak report – INFICON Explorer Portable Gas Chromatograph User Manual
Page 121
Explorer Operator Manual
119
Peak Report
The peak report displays a listing of the detected peaks. Peaks that match a library
retention time are identified and those that do not are listed as unknown.
The following information is provided for each peak:
Parameter
Description
Compound
The compound name as listed in the library. A red X
before the compound name indicates the detected
concentration has exceeded the preset alarm level.
#
The peak number in the current chromatogram.
Det
The detector used. This will be a PID or an ECD.
R.T. (sec)
The peak retention time.
Conc. (ppm)
The detected concentration as determined from the
library response value. Compounds with no response data
such as interference compounds will always be reported
as 0.0 ppm.
Area (mVS)
The integrated area. This is determined from the
integration method.
Height (mV)
The peak height.
Status
The status of the peaks.
Ratio-metric calibration may lead to an invalid
calibration and is intended for specialized applications
only. Please contact
INFICON Technical Support for
more details.