INFICON Explorer Portable Gas Chromatograph User Manual

Page 155

Advertising
background image

Explorer Operator Manual

153

Alarm Dial

sets the computer, controlling

Explorer, to dial a second computer

in the event of an alarm or fault condition.

Hang Up

terminates the modem session that is currently in progress.

SiteChart can connect to one

Explorer. In order to connect two instruments, you

must run two copies of SiteChart. You will also need a second serial port. If you
do not need to control two instruments at the same time, you may want to connect
one instrument to the computer, download the information, disconnect it and then
connect the second instrument.

Connecting Multiple Instruments

1. Start SiteChart and connect the communications cable to the first

Explorer.

2. Open the Communications menu and select

Setup

.

3. Select the serial port to which the first

Explorer is connected.

4. Ensure the first instrument is communicating with SiteChart.

5. Start another copy of SiteChart and connect the communications cable to the

second

Explorer.

6. Open the Communications menu and select

Setup

.

7. Select the serial port to which the second

Explorer is connected.

8. Ensure the second instrument is communicating with SiteChart.

SiteChart was designed to accept data from multiple instruments so the log directory
is divided into sub-directories. Each sub-directory will correspond to information
received from a single instrument. These sub-directories are created using the
instrument serial numbers. For example, if one instrument has the serial number
“123”, then all the information received from this instrument will be recorded in a
sub-directory of the data directory called “123”. The path would be C:\Program
Files\SiteChart\Data\123\filename.PID.

Advertising