Problem: There are negative dips following large positive peaks in the
chromatogram

Cause: The ECD is contaminated.

Action: Set the detector temperature to 80 C and the increase the pressure to 14
psi, or contact

INFICON.

WARNING

Do not open the instrument or attempt to remove the ECD. If the
ECD must be removed from the instrument, contact

INFICON.

Problem: Signal level too high

Cause: Components of the last analysis are still on the column.

Action: Purge the column and decrease the backflush duration to flush out the
heavier compounds. If the heavier compounds must be detected, adjust the analysis
time so these compounds have enough time to clear the analytical column, or ensure
you have selected the correct column for your application.

Cause: The ECD is contaminated.

Action: Set the detector temperature to 80 C and the increase the pressure to 14
psi, or contact

INFICON.

WARNING

Do not open the instrument or attempt to remove the ECD. If the
ECD must be removed from the instrument, contact

INFICON.