Logged data, Review logged data – INFICON Explorer Portable Gas Chromatograph User Manual

Page 85

Advertising
background image

Using

Explorer

Logged Data

The datalogger is always on, so that all analyses are logged. The datalogger will
hold approximately 40 GC log entries, which is sufficient for an 8 hour day with one
run every 10 minutes. Every GC analysis recorded in the datalogger will include a
complete peak report and chromatogram.

To view the GC log the instrument must be in the GC mode. To view the TVOC log
the instrument must be in the TVOC mode. Since TVOC logs occupy less memory,
up to 255 TVOC entries can be stored. Both GC and TVOC logs share the same
memory resource. Each successful calibration also occupies a full GC memory
space.

The status line shows the log entry number and sample date and time. The column
and the run number will be shown on a second status line. The two lines of
information will alternate.

Review Logged Data

1. Press Enter Menu, DISPLAY.

2. Select GC to review GC analyses

OR

VOC to review TVOC analyses.

3. Select the display you require.

If you are in GC mode, select PEAK, CMPD, or GRAPH.
If you are in TVOC mode, select REPORT or GRAPH.

4. Press Enter Menu to close the menu.

83