Calibration

218

Preparing a Gas Bag Sample

WARNING

Observe proper handling techniques for all gases.

NOTE: Connect the regulator to the calibration gas cylinder. You must obtain calibration

gas and a high purity regulator. Obtain a regulator matching the fitting on your
tank of calibration gas. The regulator must have a 1/4” compression fitting at the
outlet in order to use the calibration kit.

1. Tighten the regulator onto the tank with a wrench.

Do not over-tighten.

2. Attach the knurled nut on the gas bag adapter to the regulator, as shown in

Figure 10.
Finger-tighten the fitting.

1/4" to 3/16"

Union

Gas Bag

Tube Stub

1/4" Teflon Tubing

To Regulator

Outlet or

Explorer

Sample In Port

Figure 10 Gas Bag Adapter

3. Loosen the knurled nut on the reducing union of the gas bag adapter.