Explorer Operator Manual

80

4. If you are running continuously then clear the datalogger.

OR

If you are running a preset number of analyses you can clear the datalogger or
ensure there is enough free space for the number of analyses you are about to
run.

5. Press Enter Menu, then DISPLAY, VOC and GRAPH.

If you prefer, select the REPORT display option.

6. Press Enter Menu, then SETUP, MODE, and AUTO.

7. Press DELAY and use the arrow keys to enter a delay time.

This is the time between runs.
A minimum of 1 minute is required to allow the

Explorer to stabilize the

baseline.

8. Press Enter Menu to confirm the entry and close the box.

9. Press RUN and use the arrow keys to enter the number of analyses required.

If you want the instrument to sample continuously enter 0, which is the default
value.

10. Press Enter Menu to confirm the entry and close the box.

11. Press ON/OFF (the soft key) and use the arrow keys to select On and then press

Enter Menu.

12. Press Exit to close the box.

Explorer will begin a run on Column V.
Explorer will now run continuously until it has performed the preset number of

runs. If you have set the instrument to run continuously it will run until it
encounters a critical fault such as a lamp failure, low battery or carrier gas
pressure fault.