Explorer Operator Manual

62

Reviewing the Instrument Status

The status screen provides the current operating parameters. You cannot alter any
parameters in the status screen. Each parameter is listed below with a brief
description:

Parameter

Description

PID Signal

The current signal from the photoionization detector.

PID Offset

The voltage applied to keep the baseline at 10% of full scale.
The applied offset is an indication of column contamination. If
the offset is greater than 800 mV the column should be allowed
to flush until it stabilizes at a value between 200 and 800 mV.

PID UV Int

The intensity of the lamp. The intensity is measured by a
photocell and converted to a voltage. The lamp is tuned to
approximately 75% of its maximum intensity and maintained
at this intensity.

PID Status

Shows whether the lamp is on or off.

PID Temp

The operating temperature of the photoionization detector.

Column Temp

The current temperature of the selected column.

Ambient Temp

The temperature as measured by a sensor inside

Explorer.

Pressure

The column head pressure.

Active Column

The column selected for the current analysis.

Log Capacity

The amount of space remaining in the datalogger.

Battery

The

battery

voltage.